title: "VARIPULSE Bi-Directional Ablation Catheter REF D141201. The Field Catheter is... — FDA record Z-1205-2025 | Is It on Recall"
description: "Due to an observed trend of neurovascular events"
url: https://isitonrecall.com/recall/fda-device-z-1205-2025-varipulse-bi-directional-ablation-catheter-ref-d141201-the-field-catheter-is-indicated-for-use-in-ca/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# VARIPULSE Bi-Directional Ablation Catheter REF D141201. The Field Catheter is indicated for use in catheter based car... — FDA record Z-1205-2025

Reported completed

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 5, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: VARIPULSE Bi-Directional Ablation Catheter REF D141201. The Field Catheter is indicated for use in catheter based cardiac electrophysiological mapping (stimulating and recording) and, when used for TRUPULSE Generator, for the treatment of drug refractory recurrent symptomatic paroxysmal atrial fibrillation. The catheter provides location information when used with the CARTO 3 System.
- **Reason**: Due to an observed trend of neurovascular events
- **Classification**: Class I
- **Recalling firm**: Biosense Webster, Inc.
- **Quantity**: 497 total catheter - 378 US and 119 OUS
- **Agency-reported status**: Completed
- **Source record identifier**: Z-1205-2025

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1205-2025%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Product Number: D141201-12 UDI-DI code: 10846835025460 Batch Numbers: 31483574L 31483560L 31483581L 31483542L 31487313L 31483552L 31483544L 31487311L 31483550L 31483558L 31483575L 31483556L 31487312L 31483553L 31483565L 31483576L 31483547L 31487320L 31483549L 31483561L 31483548L 31483564L 31483541L 31483546L 31483562L 31483566L 31483583L 31483567L |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Worldwide distribution; US states of AR, CA, CO, KS, LA, MA, MO, NY, OH and TX. EMEA, New Zealand, Saudi Arabia, Oman, United Ecuador, Panama, South Africa, China Austria Belgium Czech Republic France Germany Hungary Ireland Israel Italy UK Sweden Norway Denmark Finland Luxembourg Netherlands Poland Portugal Slovenia Spain Switzerland Serbia United Arab Emirates Croatia Canada South Korea Japan

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Arkansas](https://isitonrecall.com/state/arkansas/)
- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Kansas](https://isitonrecall.com/state/kansas/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Massachusetts](https://isitonrecall.com/state/massachusetts/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Texas](https://isitonrecall.com/state/texas/)

## Dates in the source

- **Report date**: Mar 12, 2025
- **Recall initiation date**: Jan 5, 2025
- **Center classification date**: Feb 28, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1205-2025%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Biosense Webster, Inc. records](https://isitonrecall.com/brand/biosense-webster-inc/)