title: "ABX Pentra Calcium AS CP, Reagent 90 mL, REF A11A01954 — FDA record Z-1140-2022 | Is It on Recall"
description: "Product was shipped from a Medline warehouse to a customer at room temperature instead of refrigerated (ice). If affected product is used, the results from this"
url: https://isitonrecall.com/recall/fda-device-z-1140-2022-abx-pentra-calcium-as-cp-reagent-90-ml-ref-a11a01954/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# ABX Pentra Calcium AS CP, Reagent 90 mL, REF A11A01954 — FDA record Z-1140-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 22, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: ABX Pentra Calcium AS CP, Reagent 90 mL, REF A11A01954
- **Reason**: Product was shipped from a Medline warehouse to a customer at room temperature instead of refrigerated (ice). If affected product is used, the results from this impacted Calcium reagent will not be provided due to probable Quality Control failure. A delay in diagnosis may be experienced with repeat testing resulting in patient (customer) inconvenience, pain, and cost of the repeat blood draw and testing. All HII1220001954 product shipped/stored at a Medline branch between July 2021 and August 2021 in the identified lots were incorrectly stored and shipped. Six (6) boxes have been shipped to end users at room temperature instead of refrigerated (ice). Five (5) boxes are on hold at a Medline warehouse.
- **Classification**: Class II
- **Recalling firm**: MEDLINE INDUSTRIES, LP - Northfield
- **Quantity**: 6 boxes
- **Agency-reported status**: Terminated
- **Source record identifier**: Z-1140-2022

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1140-2022%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Model Number: HII1220001954; Lot Numbers: 215907578 (Expiration Date 09/30/2022), 215907729 (Expiration Date 02/28/2023); Manufacturer Number: A11A01954 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US Distribution to North Carolina.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [North Carolina](https://isitonrecall.com/state/north-carolina/)

## Dates in the source

- **Report date**: Jun 1, 2022
- **Recall initiation date**: Oct 22, 2021
- **Center classification date**: May 26, 2022
- **Termination date**: Jan 4, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1140-2022%22)