title: "VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-010... — FDA record Z-1070-2024 | Is It on Recall"
description: "The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity,"
url: https://isitonrecall.com/recall/fda-device-z-1070-2024-vocsn-patient-breathing-package-pediatric-active-oxygen-blue-ref-prt-01013-000-with-pouch-ref-prt-00/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Us... — FDA record Z-1070-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 9, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
- **Reason**: The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
- **Classification**: Class I
- **Recalling firm**: Ventec Life Systems, Inc.
- **Quantity**: 250 breathing circuits
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-1070-2024

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1070-2024%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Box Label - Part Numbers: PRT-01013-000 Pouch Label - Part Number: PRT-00802-001 UDI-DI Code: 00855573007594 Lot Number: 220922 Pack Slip Lot Number: 257337 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Hawaii](https://isitonrecall.com/state/hawaii/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [Washington](https://isitonrecall.com/state/washington/)

## Dates in the source

- **Report date**: Mar 6, 2024
- **Recall initiation date**: Jan 9, 2024
- **Center classification date**: Feb 23, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1070-2024%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Ventec Life Systems, Inc. records](https://isitonrecall.com/brand/ventec-life-systems-inc/)