title: "MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100) — FDA record Z-0940-2026 | Is It on Recall"
description: "Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE a"
url: https://isitonrecall.com/recall/fda-device-z-0940-2026-madsen-accuscreen-teoae-dpoae-abr-probe-p-n-8-69-41100/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100) — FDA record Z-0940-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 6, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)
- **Reason**: Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.
- **Classification**: Class II
- **Recalling firm**: PATH
- **Quantity**: Not supplied
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0940-2026

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0940-2026%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Part Number: 8-69-41100; UDI-DI: 04260223141355; Serial Numbers: 1008852, 1008888, 1008916, 1008925, 1008935, 1008945, 1008951, 1008957, 1008960, 1008962, 1008966, 1008967, 1008987, 1008990, 1009054, 1009065, 1009066, 1009131, 1009137, 1009138, 1009139, 1009145, 1009148, 1009155, 1009159, 1009161, 1009169, 1009184, 1009188, 1009191, 1009305, 1009315, 1009316, 1009319, 1009321, 1009322, 1009325, 1009328, 1009343, 1009352, 1009356, 1009357, 1009359, 1009366, 1009372, 1009375, 1009376, 1009386, 1009388, 1009390, 1009396, 1009399, 1009400, 1009404, 1009411, 1009415, 1009419, 1009426, 1009427, 1009432, 1009439, 1009442, 1009454, 1009456, 1009460, 1009463, 1009464, 1009468, 1009475, 1009481, 1009488, 1009494, 1009495, 1009497, 1009848, 1009855. |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Domestic: WI. International: Ireland.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Wisconsin](https://isitonrecall.com/state/wisconsin/)

## Dates in the source

- **Report date**: Dec 24, 2025
- **Recall initiation date**: Nov 6, 2025
- **Center classification date**: Dec 18, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.