title: "GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; Sys... — FDA record Z-0895-2025 | Is It on Recall"
description: "GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine s"
url: https://isitonrecall.com/recall/fda-device-z-0895-2025-ge-healthcare-discovery-nm-ct-670-czt-model-catalog-numbers-1-h3906ab-system-tomography-computed-emi/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission — FDA record Z-0895-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 18, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission
- **Reason**: GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- **Classification**: Class II
- **Recalling firm**: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- **Quantity**: 8 units
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0895-2025

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0895-2025%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | 1) H3906AB: UDI/DI 840682121446, System ID UX2305NU03, Serial Number CZTV29005; UDI/DI 840682121446, System ID M4166103, Serial Number CZTV29006; UDI/DI 840682121446, System ID 00243NUC22, Serial Number CZTW29008; UDI/DI 840682121446, System ID 00222NUC07, Serial Number CZTW29009; UDI/DI N/A, System ID 604875D670B, Serial Number CZTW29007; UDI/DI N/A, System ID NT6200, Serial Number AD7W38007; UDI/DI N/A, System ID NT6201, Serial Number AD7X38017; UDI/DI N/A, System ID NT6100, Serial Number AP7W37001. |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Worldwide

No US distribution state could be established from this record. Check the official notice.

## Dates in the source

- **Report date**: Jan 22, 2025
- **Recall initiation date**: Dec 18, 2024
- **Center classification date**: Jan 10, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0895-2025%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING records](https://isitonrecall.com/brand/ge-medical-systems-israel-functional-imaging/)