title: "smiths medical Medfusion Model 3500 Syringe pump — FDA record Z-0863-2024 | Is It on Recall"
description: "Medfusion syringe pumps, with software versions before v6.0.0, may have the following issues:1. Delivery During Motor Not Running High Priority Alarm, 2. Infusi"
url: https://isitonrecall.com/recall/fda-device-z-0863-2024-smiths-medical-medfusion-model-3500-syringe-pump/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# smiths medical Medfusion Model 3500 Syringe pump — FDA record Z-0863-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 19, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: smiths medical Medfusion Model 3500 Syringe pump
- **Reason**: Medfusion syringe pumps, with software versions before v6.0.0, may have the following issues:1. Delivery During Motor Not Running High Priority Alarm, 2. Infusion Restarted with Incorrect Parameters, 3. Screen Lock, 4. Interruption of Bolus or Loading Dose Delivery, 5. Pump Displays Incorrect Bolus/Loading Dose, 6. Loading/Bolus Dose Below the Minimum Recommended Rate, 7. Motor Rate Error, 8. Incorrect Recall Last Settings, 9. Corrupt Configuration, 10. Auto Lock, 11. Toolbox Configuration Loading Dose Time Values. Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software (v6.0.0) installed on your pumps.
- **Classification**: Class I
- **Recalling firm**: Smiths Medical ASD Inc.
- **Quantity**: 93177 pumps
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0863-2024

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0863-2024%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | All Medfusion Model 3500 pumps with software v3.X, v4.X, and v5.0.0 and was addressed in software v6.0.0 (2012) and all subsequent versions. |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

worldwide

No US distribution state could be established from this record. Check the official notice.

## Dates in the source

- **Report date**: Feb 7, 2024
- **Recall initiation date**: Dec 19, 2023
- **Center classification date**: Feb 1, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0863-2024%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Smiths Medical ASD Inc. records](https://isitonrecall.com/brand/smiths-medical-asd-inc/)