---
title: "Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens M... — FDA record Z-0767-2025 | Is It on Recall"
description: "Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to th"
url: https://isitonrecall.com/recall/fda-device-z-0767-2025-magnesium-flex-reagent-cartridge-360-tests-catalog-number-k3057-siemens-material-number-smn-10445158/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158; — FDA record Z-0767-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 12, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;
- **Reason**: Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out
- **Classification**: Class II
- **Recalling firm**: Siemens Healthcare Diagnostics, Inc.
- **Quantity**: 13,115 units (11,520 US, 1,595 ROW)
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0767-2025

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0767-2025%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Siemens Material Number (SMN): 10445158; Catalog Number: K3057; UDI-DI: 00842768015694; Lot numbers: 24057BB, 24064BC. |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Slovenia, Spain.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Alaska](https://isitonrecall.com/state/alaska/)
- [Alabama](https://isitonrecall.com/state/alabama/)
- [Arkansas](https://isitonrecall.com/state/arkansas/)
- [Arizona](https://isitonrecall.com/state/arizona/)
- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Connecticut](https://isitonrecall.com/state/connecticut/)
- [Delaware](https://isitonrecall.com/state/delaware/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Hawaii](https://isitonrecall.com/state/hawaii/)
- [Iowa](https://isitonrecall.com/state/iowa/)
- [Idaho](https://isitonrecall.com/state/idaho/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [Kansas](https://isitonrecall.com/state/kansas/)
- [Kentucky](https://isitonrecall.com/state/kentucky/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Massachusetts](https://isitonrecall.com/state/massachusetts/)
- [Maryland](https://isitonrecall.com/state/maryland/)
- [Maine](https://isitonrecall.com/state/maine/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [Mississippi](https://isitonrecall.com/state/mississippi/)
- [Montana](https://isitonrecall.com/state/montana/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [North Dakota](https://isitonrecall.com/state/north-dakota/)
- [Nebraska](https://isitonrecall.com/state/nebraska/)
- [New Hampshire](https://isitonrecall.com/state/new-hampshire/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New Mexico](https://isitonrecall.com/state/new-mexico/)
- [Nevada](https://isitonrecall.com/state/nevada/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Oklahoma](https://isitonrecall.com/state/oklahoma/)
- [Oregon](https://isitonrecall.com/state/oregon/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [Rhode Island](https://isitonrecall.com/state/rhode-island/)
- [South Carolina](https://isitonrecall.com/state/south-carolina/)
- [South Dakota](https://isitonrecall.com/state/south-dakota/)
- [Tennessee](https://isitonrecall.com/state/tennessee/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Utah](https://isitonrecall.com/state/utah/)
- [Virginia](https://isitonrecall.com/state/virginia/)
- [Vermont](https://isitonrecall.com/state/vermont/)
- [Washington](https://isitonrecall.com/state/washington/)
- [Wisconsin](https://isitonrecall.com/state/wisconsin/)
- [West Virginia](https://isitonrecall.com/state/west-virginia/)
- [Wyoming](https://isitonrecall.com/state/wyoming/)

## Dates in the source

- **Report date**: Jan 1, 2025
- **Recall initiation date**: Nov 12, 2024
- **Center classification date**: Dec 26, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0767-2025%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Siemens Healthcare Diagnostics, Inc. records](https://isitonrecall.com/brand/siemens-healthcare-diagnostics-inc/)
