title: "DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the pr... — FDA record Z-0677-2025 | Is It on Recall"
description: "Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISO"
url: https://isitonrecall.com/recall/fda-device-z-0677-2025-diasorin-liaison-q-s-e-t-device-plus-ref-319060-intended-for-use-in-the-preparation-of-human-stool-s/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens — FDA record Z-0677-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 7, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens
- **Reason**: Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.
- **Classification**: Class II
- **Recalling firm**: Diasorin Inc.
- **Quantity**: 21905 units
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0677-2025

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0677-2025%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | UDI/DI 80567713190605F, Lot Numbers: 225084, 232094, 234114, 259144, 221214, 252244, 230094, 224124, 205144, 228174, 228224, 223274, 219104, 217134, 233154, 236174, 224244, 232294, 229224, 251234, 221294, 210204, 223244, 222314 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Worldwide

No US distribution state could be established from this record. Check the official notice.

## Dates in the source

- **Report date**: Dec 18, 2024
- **Recall initiation date**: Nov 7, 2024
- **Center classification date**: Dec 12, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0677-2025%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Diasorin Inc. records](https://isitonrecall.com/brand/diasorin-inc/)