title: "smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model... — FDA record Z-0466-2025 | Is It on Recall"
description: "The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or mor"
url: https://isitonrecall.com/recall/fda-device-z-0466-2025-smiths-medical-cadd-solis-li-ion-rechargeable-battery-packs-ac-adapter-model-number-ref-21-0270-25-f/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the... — FDA record Z-0466-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 26, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps
- **Reason**: The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.
- **Classification**: Class II
- **Recalling firm**: Smiths Medical ASD, Inc.
- **Quantity**: 156,400 AC adapters
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0466-2025

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0466-2025%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Model 21-0270-25, UDI/DI 15019517126844, All lots |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Worldwide distribution.

No US distribution state could be established from this record. Check the official notice.

## Dates in the source

- **Report date**: Nov 27, 2024
- **Recall initiation date**: Jun 26, 2024
- **Center classification date**: Nov 15, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0466-2025%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Smiths Medical ASD Inc. records](https://isitonrecall.com/brand/smiths-medical-asd-inc/)