title: "Simplexa Direct Amplification Disc Kit, Product Kit Code MOL1455 — FDA record Z-0457-2022 | Is It on Recall"
description: "A subset of affected discs are at risk of leakage when it is used more than one time with Simplexa COVID-19 Direct and/or Simplexa Flu A/B & RSV Direct Gen II."
url: https://isitonrecall.com/recall/fda-device-z-0457-2022-simplexa-direct-amplification-disc-kit-product-kit-code-mol1455/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Simplexa Direct Amplification Disc Kit, Product Kit Code MOL1455 — FDA record Z-0457-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 19, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Simplexa Direct Amplification Disc Kit, Product Kit Code MOL1455
- **Reason**: A subset of affected discs are at risk of leakage when it is used more than one time with Simplexa COVID-19 Direct and/or Simplexa Flu A/B & RSV Direct Gen II. Leakage from a well that had a positive sample may cause further cross-contamination in areas of the lab if not properly cleaned.
- **Classification**: Class II
- **Recalling firm**: DiaSorin Molecular LLC
- **Quantity**: 29495 kits
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0457-2022

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0457-2022%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Kit Product Code MOL1455 (UDI 01-30816101025092) contains 3 discs with Disc Product Code MOL1452. Below lists the MOL1455 Kit Lot numbers (and MOL1452 Disc Lot Numbers): 12410N (12152N), 12411N (12147N), 12413N (12149N), 12477N (12150N), 12602N (11548NA), 12603N (12144N), 12721N (12148N), 12889N (12151N), 13369N (12817N), 13370N (13151N), 13434N (12382N), 13652N (12835N), and 13242N (12154N)\*. \*Kit Lot #13242N (Disc Lot #12154N) was shipped to Europe and not within the US. |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US: AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV; and Worldwide: Kuwait, Australia, Israel, Italy, Canada, Chile, Puerto Rico, United Arab Emirates.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Arizona](https://isitonrecall.com/state/arizona/)
- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Connecticut](https://isitonrecall.com/state/connecticut/)
- [District of Columbia](https://isitonrecall.com/state/district-of-columbia/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Hawaii](https://isitonrecall.com/state/hawaii/)
- [Iowa](https://isitonrecall.com/state/iowa/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [Kansas](https://isitonrecall.com/state/kansas/)
- [Kentucky](https://isitonrecall.com/state/kentucky/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Massachusetts](https://isitonrecall.com/state/massachusetts/)
- [Maryland](https://isitonrecall.com/state/maryland/)
- [Maine](https://isitonrecall.com/state/maine/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [North Dakota](https://isitonrecall.com/state/north-dakota/)
- [Nebraska](https://isitonrecall.com/state/nebraska/)
- [New Hampshire](https://isitonrecall.com/state/new-hampshire/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New Mexico](https://isitonrecall.com/state/new-mexico/)
- [Nevada](https://isitonrecall.com/state/nevada/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [Rhode Island](https://isitonrecall.com/state/rhode-island/)
- [South Carolina](https://isitonrecall.com/state/south-carolina/)
- [South Dakota](https://isitonrecall.com/state/south-dakota/)
- [Tennessee](https://isitonrecall.com/state/tennessee/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Virginia](https://isitonrecall.com/state/virginia/)
- [Vermont](https://isitonrecall.com/state/vermont/)
- [Washington](https://isitonrecall.com/state/washington/)
- [Wisconsin](https://isitonrecall.com/state/wisconsin/)
- [West Virginia](https://isitonrecall.com/state/west-virginia/)

## Dates in the source

- **Report date**: Jan 12, 2022
- **Recall initiation date**: Nov 19, 2021
- **Center classification date**: Jan 5, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0457-2022%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [DiaSorin Molecular LLC records](https://isitonrecall.com/brand/diasorin-molecular-llc/)