title: "Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number ... — FDA record Z-0452-2022 | Is It on Recall"
description: "The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicate"
url: https://isitonrecall.com/recall/fda-device-z-0452-2022-biomeme-sars-cov-2-real-time-rt-pcr-test-instructions-for-use-part-number-100002824-subcomponent-of/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go... — FDA record Z-0452-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 20, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564
- **Reason**: The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).
- **Classification**: Class II
- **Recalling firm**: Biomeme, Inc.
- **Quantity**: 11,607
- **Agency-reported status**: Terminated
- **Source record identifier**: Z-0452-2022

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0452-2022%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | All lots. Last lot of Kit distributed with the old IFU is 20210914. |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Domestic distribution to AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK, PA, TX, VA. Foreign distribution to Ghana, Dominican Republic, and Singapore.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Arizona](https://isitonrecall.com/state/arizona/)
- [California](https://isitonrecall.com/state/california/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Hawaii](https://isitonrecall.com/state/hawaii/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [Kentucky](https://isitonrecall.com/state/kentucky/)
- [Maryland](https://isitonrecall.com/state/maryland/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Oklahoma](https://isitonrecall.com/state/oklahoma/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Virginia](https://isitonrecall.com/state/virginia/)

## Dates in the source

- **Report date**: Jan 12, 2022
- **Recall initiation date**: Oct 20, 2021
- **Center classification date**: Jan 4, 2022
- **Termination date**: Mar 5, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0452-2022%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Biomeme, Inc. records](https://isitonrecall.com/brand/biomeme-inc/)