# SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FI... — FDA record Z-0372-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 5, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL¿ End-to-End Capillary Tubes, 30 ¿L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910
- **Reason**: All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.
- **Classification**: Class II
- **Recalling firm**: ASP Global, LLC. dba Anatomy Supply Partners, LLC.
- **Quantity**: 4,272,858
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0372-2026

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0372-2026%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Model/UDI-DI/Lots: 07 6011/10643351000014/23H4110, 23I4017, 23I4051, 23J4031, 23K4058, 23K4059, 23L4082, 24A4103, 24B4186, 24B4187, 24D4015, 24E4138, 24E4139, 24H4121, 24I4317, 24I4328, 24J4041, 24J4137, 24J4138, 24J4233, 24J4316, 24J4348; 07 6013/10643351000021/23H4111, 23I4018, 24B4188, 24H4125, 25B4099; 07 7051/10643351000038/23H4157, 23I4020, 23I4374, 23J4033, 23K4060, 23L4083, 24A4104, 24B4190, 24D4016, 24E4140, 24F4053, 24I4321, 24J4043, 24J4139, 24J4237, 24L4001, 25B4107; 07 7052/10643351000045/23H4113, 23I4021, 23J4034, 24A4105, 24D4017, 24F4052, 24H4126, 24I4323, 24J4042, 24J4140, 24J4338, 24J4355, 25B4108; 07 7053/10643351000052/23I4022, 23J4035, 23L4084, 24D4018, 24E4132, 24F4192, 25B4106; 07 7056/10643351000069/24A4106, 24B4191, 24B4192, 24E4133, 24I4315, 24J4136, 24J4347; 07 7058/10643351000076/23G4430, 23G4168. Manufactured between July 31, 2023 and February 28, 2025. |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Alaska](https://isitonrecall.com/state/alaska/)
- [Alabama](https://isitonrecall.com/state/alabama/)
- [Arkansas](https://isitonrecall.com/state/arkansas/)
- [Arizona](https://isitonrecall.com/state/arizona/)
- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Connecticut](https://isitonrecall.com/state/connecticut/)
- [Delaware](https://isitonrecall.com/state/delaware/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Hawaii](https://isitonrecall.com/state/hawaii/)
- [Iowa](https://isitonrecall.com/state/iowa/)
- [Idaho](https://isitonrecall.com/state/idaho/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [Kansas](https://isitonrecall.com/state/kansas/)
- [Kentucky](https://isitonrecall.com/state/kentucky/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Massachusetts](https://isitonrecall.com/state/massachusetts/)
- [Maryland](https://isitonrecall.com/state/maryland/)
- [Maine](https://isitonrecall.com/state/maine/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [Mississippi](https://isitonrecall.com/state/mississippi/)
- [Montana](https://isitonrecall.com/state/montana/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [North Dakota](https://isitonrecall.com/state/north-dakota/)
- [Nebraska](https://isitonrecall.com/state/nebraska/)
- [New Hampshire](https://isitonrecall.com/state/new-hampshire/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New Mexico](https://isitonrecall.com/state/new-mexico/)
- [Nevada](https://isitonrecall.com/state/nevada/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Ohio](https://isitonrecall.com/state/ohio/)
- [Oklahoma](https://isitonrecall.com/state/oklahoma/)
- [Oregon](https://isitonrecall.com/state/oregon/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [Rhode Island](https://isitonrecall.com/state/rhode-island/)
- [South Carolina](https://isitonrecall.com/state/south-carolina/)
- [South Dakota](https://isitonrecall.com/state/south-dakota/)
- [Tennessee](https://isitonrecall.com/state/tennessee/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Utah](https://isitonrecall.com/state/utah/)
- [Virginia](https://isitonrecall.com/state/virginia/)
- [Vermont](https://isitonrecall.com/state/vermont/)
- [Washington](https://isitonrecall.com/state/washington/)
- [Wisconsin](https://isitonrecall.com/state/wisconsin/)
- [West Virginia](https://isitonrecall.com/state/west-virginia/)
- [Wyoming](https://isitonrecall.com/state/wyoming/)

## Dates in the source

- **Report date**: Nov 5, 2025
- **Recall initiation date**: May 5, 2025
- **Center classification date**: Oct 27, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0372-2026%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [ASP Global, LLC. dba Anatomy Supply Partners, LLC. records](https://isitonrecall.com/brand/asp-global-llc-dba-anatomy-supply-partners-llc/)