title: "Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite ... — FDA record Z-0274-2022 | Is It on Recall"
description: "Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and cali"
url: https://isitonrecall.com/recall/fda-device-z-0274-2022-freelite-human-lambda-free-kit-for-use-on-the-spaplus-ref-lk018-s-freelite-human-lambda-free-kit-for/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SP... — FDA record Z-0274-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 15, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S
- **Reason**: Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6 activity reaches the optical ceiling of the SPA+ analyzer.
- **Classification**: Class II
- **Recalling firm**: The Binding Site Group, Ltd.
- **Quantity**: 2,254 kits
- **Agency-reported status**: Terminated
- **Source record identifier**: Z-0274-2022

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0274-2022%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Model Number: LK018.S UDI Code: 05051700006477 Lot Number: 472776 Model Number:LK018.10S UDI Code: 05051700012201 Lot Number: 472664 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

U.S.: CA, CO, FL, IL, IN, KY, LA, MI, NY, OK, SD, TX and VA. O.U.S.: Canada

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [Kentucky](https://isitonrecall.com/state/kentucky/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Oklahoma](https://isitonrecall.com/state/oklahoma/)
- [South Dakota](https://isitonrecall.com/state/south-dakota/)
- [Texas](https://isitonrecall.com/state/texas/)
- [Virginia](https://isitonrecall.com/state/virginia/)

## Dates in the source

- **Report date**: Dec 1, 2021
- **Recall initiation date**: Oct 15, 2021
- **Center classification date**: Nov 19, 2021
- **Termination date**: Feb 4, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0274-2022%22)