title: "EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12... — FDA record Z-0225-2025 | Is It on Recall"
description: "Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surge"
url: https://isitonrecall.com/recall/fda-device-z-0225-2025-ezdilate-wire-guided-balloon-indicated-for-use-in-adult-and-adolescent-12-years-populations-to-endos/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically di... — FDA record Z-0225-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 13, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy. Model/Catalog Number: BD-410X-1055
- **Reason**: Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon, including the Shelf Box with Front Box Label and Circular Star Label and the balloon Pouch
- **Classification**: Class II
- **Recalling firm**: Olympus Corporation of the Americas
- **Quantity**: 209 units
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0225-2025

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0225-2025%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | UDI-DI: 821925033238 Lot: 408987 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Worldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA, TN, VA, WI; and Foreign OUS to countries of: CA, DE, IN, SG, Latin America.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [California](https://isitonrecall.com/state/california/)
- [Delaware](https://isitonrecall.com/state/delaware/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Indiana](https://isitonrecall.com/state/indiana/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [Tennessee](https://isitonrecall.com/state/tennessee/)
- [Virginia](https://isitonrecall.com/state/virginia/)
- [Wisconsin](https://isitonrecall.com/state/wisconsin/)

## Dates in the source

- **Report date**: Oct 30, 2024
- **Recall initiation date**: Sep 13, 2024
- **Center classification date**: Oct 24, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0225-2025%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Olympus Corporation of the Americas records](https://isitonrecall.com/brand/olympus-corporation-of-the-americas/)