title: "GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with softwar... — FDA record Z-0213-2026 | Is It on Recall"
description: "Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring"
url: https://isitonrecall.com/recall/fda-device-z-0213-2026-ge-healthcare-carescape-central-station-cscs-v3-mas800-desktop-with-software-version-v3-0-5-part-num/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04... — FDA record Z-0213-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 5, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube
- **Reason**: Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
- **Classification**: Class II
- **Recalling firm**: Wipro GE Healthcare Private Ltd.
- **Quantity**: 173 units
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0213-2026

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0213-2026%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | GTIN 00195278343512, Serial Numbers: STU25210418SA, STU25210417SA, STU25210419SA, STU25250440SA, STU25250439SA, STU25260447SA, STU25210407SA, STU25210406SA, STU24200003SA, STU24200006SA, STU24200005SA, STU24200004SA, STU23240015SA, STU24240019SA, STU24240020SA, STU24240007SA, STU24240022SA, STU24240021SA, STU24330020SA, STU24240014SA, STU24240013SA, STU24330019SA, STU24330022SA, STU24330016SA, STU24240005SA, STU24240008SA, STU24240012SA, STU24330018SA, STU24330021SA, STU24330017SA, STU23160011SA, STU23240007SA, STU24310001SA, STU25210433SA, STU25210432SA, STU25200390SA, STU25200392SA, STU25200391SA, STU25200399SA, STU25200397SA, STU25200395SA, STU25200393SA, STU25200398SA, STU25200396SA, STU25250438SA, STU25270475SA, STU25270456SA, STU25270468SA, STU25270458SA, STU25270466SA, STU25270476SA, STU25270469SA, STU25270473SA, STU25270459SA, STU25270474SA, STU25270467SA, STU25210430SA, STU25210429SA, STU25180343SA, STU25190347SA, STU25190348SA, STU24200007SA, STU25190360SA, STU25240436SA, STU25190361SA, STU25190362SA, STU25190363SA, STU25240437SA, STU25190359SA, STU24510027SA, STU24510032SA, STU24510010SA, STU24510020SA, STU24510017SA, STU24510026SA, STU24510018SA, STU24510024SA, STU24510025SA, STU24510031SA, STU24510011SA, STU24510029SA, STU24510023SA, STU24510030SA, STU24510021SA, STU24510019SA, STU24510022SA, STU25190350SA, STU25200388SA, STU25200389SA, STU25210412SA, STU25190349SA, STU25190381SA, STU25190376SA, STU25190383SA, STU25190374SA, STU25190367SA, STU25190372SA, STU25190379SA, STU25190368SA, STU25190352SA, STU25190358SA, STU25190370SA, STU25190366SA, STU25190355SA, STU25190354SA, STU25190382SA, STU25190380SA, STU25190371SA, STU25190377SA, STU25190353SA, STU25190351SA, STU25190364SA, STU25190357SA, STU25190384SA, STU25190369SA, STU25190373SA, STU25190356SA, STU25190365SA, STU25190378SA, STU25190385SA, STU25210410SA, STU25210411SA, STU25210409SA, STU25230435SA, STU25180346SA, STU25180344SA, STU25180345SA, STU25210413SA, STU25210414SA, STU25210423SA, STU25210431SA, STU25210415SA, STU25270453SA, STU25260450SA, STU25260451SA, STU25290488SA, STU25210408SA, STU25210400SA, STU25270454SA, STU25210404SA, STU25210421SA, STU25270464SA, STU25210420SA, STU25270462SA, STU25210425SA, STU25270461SA, STU25210426SA, STU25210428SA, STU25210427SA, STU25270463SA, STU25270460SA, STU24460058SA, STU24460059SA, STU24460071SA, STU24460060SA, STU25200387SA, STU25210401SA, STU25200386SA, STU25210424SA, STU25270457SA, STU25270465SA, STU25300492SA, STU25210402SA, STU25210403SA, STU25250446SA, STU25210422SA, STU25210405SA, STU25260449SA, STU25290481SA, STU25290482SA, STU25290483SA, STU25290484SA, STU25210416SA. |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

Worldwide distribution.

No US distribution state could be established from this record. Check the official notice.

## Dates in the source

- **Report date**: Oct 29, 2025
- **Recall initiation date**: Sep 5, 2025
- **Center classification date**: Oct 21, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0213-2026%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Wipro GE Healthcare Private Ltd. records](https://isitonrecall.com/brand/wipro-ge-healthcare-private-ltd/)