---
title: "Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI... — FDA record Z-0198-2023 | Is It on Recall"
description: "After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal conf"
url: https://isitonrecall.com/recall/fda-device-z-0198-2023-canon-aplio-ultrasound-system-model-aplio-i900-tus-ai900-aplio-i800-tus-ai800-aplio-i700-tus-ai700-a/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CU... — FDA record Z-0198-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 29, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450
- **Reason**: After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.
- **Classification**: Class II
- **Recalling firm**: Canon Medical System, USA, INC.
- **Quantity**: 65
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0198-2023

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0198-2023%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Model/UDI-DI, with software: Aplio i900 TUS-AI900/ 04987670102523, Aplio i800 TUS-AI800/ 04987670102516, Aplio i700 TUS-AI700/ 04987670102509, With System Version:V4.6/V5.0/V5.1, SP0005 to SP1014, or V6.0/V6.5 SP0000 to SP1009; Aplio a550 CUS-AA550/ 04987670103353, Aplio a450 CUS-AA450/ 04987670103346, With System Version: V3.0, SP0005 to SP1014, or V4.0/V4.5, SP0000 to SP1009 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA, OK, ME, KY, NY, PA, GA, FL

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Arkansas](https://isitonrecall.com/state/arkansas/)
- [California](https://isitonrecall.com/state/california/)
- [Colorado](https://isitonrecall.com/state/colorado/)
- [Florida](https://isitonrecall.com/state/florida/)
- [Georgia](https://isitonrecall.com/state/georgia/)
- [Iowa](https://isitonrecall.com/state/iowa/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Kansas](https://isitonrecall.com/state/kansas/)
- [Kentucky](https://isitonrecall.com/state/kentucky/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Maine](https://isitonrecall.com/state/maine/)
- [Michigan](https://isitonrecall.com/state/michigan/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [Missouri](https://isitonrecall.com/state/missouri/)
- [Montana](https://isitonrecall.com/state/montana/)
- [North Carolina](https://isitonrecall.com/state/north-carolina/)
- [New York](https://isitonrecall.com/state/new-york/)
- [Oklahoma](https://isitonrecall.com/state/oklahoma/)
- [Pennsylvania](https://isitonrecall.com/state/pennsylvania/)
- [South Carolina](https://isitonrecall.com/state/south-carolina/)
- [Tennessee](https://isitonrecall.com/state/tennessee/)
- [Washington](https://isitonrecall.com/state/washington/)
- [West Virginia](https://isitonrecall.com/state/west-virginia/)

## Dates in the source

- **Report date**: Nov 16, 2022
- **Recall initiation date**: Sep 29, 2022
- **Center classification date**: Nov 7, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0198-2023%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Canon Medical System, USA, INC. records](https://isitonrecall.com/brand/canon-medical-system-usa-inc/)
