title: "Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ... — FDA record Z-0170-2024 | Is It on Recall"
description: "Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm m"
url: https://isitonrecall.com/recall/fda-device-z-0170-2024-senhance-surgical-system-which-includes-manipulator-arm-isu-configuration-catalog-number-x9007708-in/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent ... — FDA record Z-0170-2024

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 15, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system
- **Reason**: Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.
- **Classification**: Class I
- **Recalling firm**: Asensus Surgical Inc
- **Quantity**: 21 units
- **Agency-reported status**: Terminated
- **Source record identifier**: Z-0170-2024

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0170-2024%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | UDI/DI 00815440022075, Serial Numbers: SN 00012, SN 00016, SN 00007, SN 00017, SN 00035, SN 00018, SN 00042, SN 00008, SN 00031, SN 00025, SN 00011, SN 00028, SN 00038, SN 00029, SN 00036, SN 00025, SN 00024, SN 00006, SN 00019, SN 00040, SN 00041, loaded with Software (SW) version 2.7.4 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Florida](https://isitonrecall.com/state/florida/)
- [Illinois](https://isitonrecall.com/state/illinois/)
- [Louisiana](https://isitonrecall.com/state/louisiana/)
- [Minnesota](https://isitonrecall.com/state/minnesota/)
- [New Jersey](https://isitonrecall.com/state/new-jersey/)

## Dates in the source

- **Report date**: Nov 8, 2023
- **Recall initiation date**: Sep 15, 2023
- **Center classification date**: Oct 30, 2023
- **Termination date**: Sep 26, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0170-2024%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [Asensus Surgical Inc records](https://isitonrecall.com/brand/asensus-surgical-inc/)