title: "LFS Flat screen arm system with surgical lamp for Models Aurora series, Auror... — FDA record Z-0071-2024 | Is It on Recall"
description: "The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate b"
url: https://isitonrecall.com/recall/fda-device-z-0071-2024-lfs-flat-screen-arm-system-with-surgical-lamp-for-models-aurora-series-aurora-2-series-aurora-3-seri/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Recall archive](https://isitonrecall.com/archives/) / FDA device enforcement

# LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 se... — FDA record Z-0071-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 23, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## What the record says

- **Product**: LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23, LFSST2323TV, LFSLFSST2323, LDFSLFSLED7TV5, LDFSLFSLED7TV55, LFSLED55, LFSLFSST23, LFSST2923TV, LFSST29TV, LFSST29, LFSST19WC, LFSST2923, LFSLFSST-1, LFSST292323, LFSST1919WC, LFS ST2319WC
- **Reason**: The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.
- **Classification**: Class II
- **Recalling firm**: DKK Dai-Ichi Shomei Co., Ltd.
- **Quantity**: 256 units
- **Agency-reported status**: Ongoing
- **Source record identifier**: Z-0071-2024

[Open the official record](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0071-2024%22)

[What the recall classes mean](https://isitonrecall.com/guides/recall-classes/). Classification is an agency assessment, not individual medical advice.

## Lot and product identifiers

| Label field | Details supplied by the agency |
| --- | --- |
| Lot, code or applicability | Serial Numbers: 06087323 - 06087352, 06087353 - 06087354, 06081292 - 06081293, 06087448 - 06087449, 06087450 - 06087469, 06097595 - 06097596, 06107716 - 06107718, 06107719 - 06107728, 06107730, 06107731 - 06107732, 06111348 - 06111349, 06117855 - 06117856, 06117857 - 06117866, 06117868 - 06117868, 06111350 - 06111351, 06117949 - 06117949, 06117954 - 06117973, 06118031 - 06118032, 06128116 - 06128131, 06128172 - 06128176, 06128177 - 06128186, 06128187 - 06128188, 07010004, 07010002, 07010005 - 07010006, 07028436 - 07028441, 07030080 - 07030081, 07048491 - 07048495, 07058506 - 07058507, 07058508 - 07058512, 07068602 - 07068604, 07068605 - 07068607, 07068611 - 07068622, 07068609 - 07068610, 07068667 - 07068670, 07061479 - 07061484, 07078679 - 07078680, 07071485 - 07071488, 07071501, 07088741 - 07088742, 07081124, 07088766, 07098769 - 07098773, 07098767 - 07098768, 07098774 - 07098775, 07108885 - 07108889, 07108883 - 07108884, 07108876 - 07108882, 07108890 - 07108891, 07101533 - 07101534, 07129027 - 07129028, RA000182, RA000184, 08029174 - 08029183, 08029187 - 08029188, 08029189 - 08029190, 09111694, 10011776, 10011777, 10011778, 10011779, 10011780, 10011781, 10011782, 10051792, 10051794, 10051796, 10051798, 10051800, 10051802, 10051804, 100610044, 100610045, 100610046, 100610047, 100610048, 100610049, 100610050, 100610051, 100610052, 100610053, 101010180, 101010182, 101010184, 101010186, 101010188, 101010190, 101010192, 101210252, BM005435 - BM005435, BM006121 - BM006121, BM008121 - BM008127 |

Match all identifying details in the current notice. A brand name alone is not enough.

## Distribution

US States: MI

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

- [Michigan](https://isitonrecall.com/state/michigan/)

## Dates in the source

- **Report date**: Oct 11, 2023
- **Recall initiation date**: Aug 23, 2023
- **Center classification date**: Oct 5, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

## What to do next

Read the official notice for the action that applies to your exact product.

[Read the current instructions](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0071-2024%22)

## Related records

[Medical devices records](https://isitonrecall.com/category/medical-devices/) [DKK Dai-Ichi Shomei Co., Ltd. records](https://isitonrecall.com/brand/dkk-dai-ichi-shomei-co-ltd/)