title: "NOXBOX LTD Recalls: Agency-Reported Status and History (Oct 1, 2026) | Is It on Recall"
description: "Search NOXBOX LTD recall records, lots, firms and dates. Reported status can be incomplete; use the linked agency notice for current instructions."
url: https://isitonrecall.com/brand/noxbox-ltd/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Home](https://isitonrecall.com/) / Brand archive

# NOXBOX LTD recall records

**4** reported ongoing

5 records in this archive. Latest source date Aug 19, 2026.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## Records and reported status

Aug 19, 2026 FDA device enforcement

### [NOxBOXi Nitric Oxide Delivery System.](https://isitonrecall.com/recall/fda-device-z-3165-2026-noxboxi-nitric-oxide-delivery-system/)

Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.

Class II Reported ongoing

Distribution: US Nationwide distribution in the state of Georgia.

[Official source for NOxBOXi Nitric Oxide Delivery System.](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-3165-2026%22)

Sep 9, 2025 FDA device enforcement

### [NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I](https://isitonrecall.com/recall/fda-device-z-0028-2026-noxboxi-nitric-oxide-delivery-system-model-numbers-noxbox-i/)

Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV ...

Class I Reported ongoing

Distribution: Domestic: TN

[Official source for NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0028-2026%22)

Sep 9, 2025 FDA device enforcement

### [NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I](https://isitonrecall.com/recall/fda-device-z-0029-2026-noxboxi-nitric-oxide-delivery-system-model-numbers-noxbox-i/)

The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.

Class I Reported ongoing

Distribution: Domestic: TN;

[Official source for NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0029-2026%22)

Jul 19, 2023 FDA device enforcement

### [NOxBOXi Nitric Oxide Delivery System](https://isitonrecall.com/recall/fda-device-z-2544-2025-noxboxi-nitric-oxide-delivery-system/)

Replacement of the internal Sample Pump due to high failure rates.

Class II Reported ongoing

Distribution: US Nationwide distribution in the state of TN.

[Official source for NOxBOXi Nitric Oxide Delivery System](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2544-2025%22)

May 3, 2023 FDA device enforcement

### [NOxBOXi Nitric Oxide Delivery System, NOXBOX-I, REQNOXBOX-I, Nitric Oxide Delivery Device](https://isitonrecall.com/recall/fda-device-z-1667-2023-noxboxi-nitric-oxide-delivery-system-noxbox-i-reqnoxbox-i-nitric-oxide-delivery-device/)

Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a worst case scenario, ...

Class I Reported terminated

Distribution: Worldwide distribution - US Nationwide and the countries of Bahrain, Bolivia, Brazil, Ecuador, Finland, France, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Kuwait, Mala...

[Official source for NOxBOXi Nitric Oxide Delivery System, NOXBOX-I, REQNOXBOX-I, Nitric Oxide Delivery Device](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1667-2023%22)

## Reasons in these listed records

- Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.
- Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or convention...
- The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
- Replacement of the internal Sample Pump due to high failure rates.

These examples come from this result page and are not a ranking of every recall reason.

## Checking NOXBOX LTD

### Is there a recall on noxbox ltd right now?

This archive contains 5 records, including 4 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

### How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. [Use the label-check guide](https://isitonrecall.com/guides/read-lot-codes/) if you need help locating an identifier.

### Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

### What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. [Check source coverage](https://isitonrecall.com/sources/) and the agency's current notices.

### Where can I find older records?

Use [monthly archives](https://isitonrecall.com/archives/) or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

## Keep checking