# Nobel Biocare recall records

**2** reported ongoing

2 records in this archive. Latest source date Jan 10, 2025.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## Records and reported status

Jan 10, 2025 FDA device enforcement

### [. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intende...](https://isitonrecall.com/recall/fda-device-z-1101-2025-nobel-biocare-n1-tiultra-tcc-rp-4-0x9mm-nobel-biocare-n1-tiultra-tcc-rp-4-0x11mm-nobel-biocare-n1-t/)

Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.

Class II Reported ongoing

Distribution: U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.

[Official source for . Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Bioc...](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1101-2025%22)

Jul 25, 2022 FDA device enforcement

### [Nobel Biocare N1 TiUltra TCC NP 3.5x13mm](https://isitonrecall.com/recall/fda-device-z-1565-2022-nobel-biocare-n1-tiultra-tcc-np-3-5x13mm/)

Due to the presence of a burr which may cause an aspiration risk to patient during surgical procedure.

Class II Reported ongoing

Distribution: U.S Nationwide.: AZ, CA, CT, FL, HI, NC, NY, OR, PA, TX, UT, and WI O.U.S.: Austria, Belgium, Canada, Germany, Spain, France, United Kingdom, Hungary, Iceland, Italy, Netherland...

[Official source for Nobel Biocare N1 TiUltra TCC NP 3.5x13mm](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1565-2022%22)

## Reasons in these listed records

- Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.
- Due to the presence of a burr which may cause an aspiration risk to patient during surgical procedure.

These examples come from this result page and are not a ranking of every recall reason.

## Checking Nobel Biocare

### Is there a recall on nobel biocare right now?

This archive contains 2 records, including 2 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

### How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. [Use the label-check guide](https://isitonrecall.com/guides/read-lot-codes/) if you need help locating an identifier.

### Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

### What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. [Check source coverage](https://isitonrecall.com/sources/) and the agency's current notices.

### Where can I find older records?

Use [monthly archives](https://isitonrecall.com/archives/) or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

## Keep checking