title: "Cardiac Assist, Inc Recalls: Agency-Reported Status and History (Oct 1, 2026) | Is It on Recall"
description: "Search Cardiac Assist, Inc recall records, lots, firms and dates. Reported status can be incomplete; use the linked agency notice for current instructions."
url: https://isitonrecall.com/brand/cardiac-assist-inc/
updated: 2026-10-01
site: "Is It on Recall"
language: en
---

[Home](https://isitonrecall.com/) / Brand archive

# Cardiac Assist, Inc recall records

**2** reported ongoing

4 records in this archive. Latest source date Jun 30, 2023.

FDA enforcement status can lag behind the recall lifecycle. An empty search or a reported termination is not a personal product-safety determination.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; [check FSIS directly](https://www.fsis.usda.gov/recalls). [Source coverage and dates](https://isitonrecall.com/sources/).

## Records and reported status

Jun 30, 2023 FDA device enforcement

### [LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray part number: 5800-1000. Note: The 5800-1000 model number is an individually packaged co...](https://isitonrecall.com/recall/fda-device-z-2465-2023-lifesparc-pump-part-number-5800-0000-contained-within-tandemlife-lifesparc-priming-tray-part-number/)

Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,

Class II Reported ongoing

Distribution: US Nationwide distribution in the states of IL, FL.

[Official source for LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray part nu...](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2465-2023%22)

Dec 5, 2022 FDA device enforcement

### [SHIP KIT, LIFESPARC, CONTROLLER AND ACCESSORIES - 5900-0000; SHIP KIT, LIFESPARC, CONTROLLER ONLY - 5900-0001. Component of LifeSPARC System, used to pump the blood through an e...](https://isitonrecall.com/recall/fda-device-z-0926-2023-ship-kit-lifesparc-controller-and-accessories-5900-0000-ship-kit-lifesparc-controller-only-5900-0001/)

A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The Critical Failure was first addressed in the firm's recall initiated July 21, 2022. During the Critical Failure, the softw...

Class I Reported terminated

Distribution: US Nationwide distribution.

[Official source for SHIP KIT, LIFESPARC, CONTROLLER AND ACCESSORIES - 5900-0000; SHIP KIT, LIFESPARC, CONTROLLER ONLY...](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0926-2023%22)

Jul 21, 2022 FDA device enforcement

### [TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model 5900-0000; Controller Only Model 5900-0001. For circulatory or cardiopulmonary bypass.](https://isitonrecall.com/recall/fda-device-z-1763-2022-tandem-life-livanova-lifesparc-controller-controller-and-accessories-model-5900-0000-controller-only/)

Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.

Class I Reported terminated

Distribution: US Nationwide distribution.

[Official source for TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model 5900-0000; Controller...](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1763-2022%22)

Oct 14, 2021 FDA device enforcement

### [SHIP KIT, PROCEDURE KIT, PROTEKDUO LS - DL31 for temporary Cardiopulmonary bypass. Including the following components: a) 31 Fr ProtekDuo Veno-Venous Cannula Set b) LifeSPARC Pu...](https://isitonrecall.com/recall/fda-device-z-0311-2022-ship-kit-procedure-kit-protekduo-ls-dl31-for-temporary-cardiopulmonary-bypass-including-the-followin/)

Top label of the outer package of the collection kit Item 5820-3118 showed expiration date of 2022/05/01 (correct expiration date) while side label showed expiration date of 2021/05/01 (incorrect expiration date)

Class III Reported ongoing

Distribution: US Distribution to states of: DC, MA, SC, and VA.

[Official source for SHIP KIT, PROCEDURE KIT, PROTEKDUO LS - DL31 for temporary Cardiopulmonary bypass. Including the ...](https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0311-2022%22)

## Reasons in these listed records

- Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,
- A software update (v1.1.5) has been developed to address the issue of Critical Failure which can occur on the LifeSPARC Controller. The Critical Failure was first addressed in the firm's recall initiated July 21, 2022. During the Critical Failure, the software freezes or crash...
- Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.
- Top label of the outer package of the collection kit Item 5820-3118 showed expiration date of 2022/05/01 (correct expiration date) while side label showed expiration date of 2021/05/01 (incorrect expiration date)

These examples come from this result page and are not a ranking of every recall reason.

## Checking Cardiac Assist, Inc

### Is there a recall on cardiac assist, inc right now?

This archive contains 4 records, including 2 with agency-reported ongoing status. The status field can be incomplete or stale. Open the current official notice and match your exact product.

### How do I check my package or product?

Compare the product, firm and lot, date, model or serial identifiers with the notice. [Use the label-check guide](https://isitonrecall.com/guides/read-lot-codes/) if you need help locating an identifier.

### Does a matching name mean every item is included?

A name is a search match. The agency notice sets the affected products and identifiers. Do not assume that every item with the same name or brand belongs to the recall.

### What if the search finds no matching record?

This search has source and keyword limits. Missing records or a missing lot code do not establish that a product is unaffected. [Check source coverage](https://isitonrecall.com/sources/) and the agency's current notices.

### Where can I find older records?

Use [monthly archives](https://isitonrecall.com/archives/) or the local search. The initial import covers records dated from October 2021 where available; each source's actual first and latest dates appear on the coverage page.

## Keep checking

[Search Cardiac Assist, Inc](https://isitonrecall.com/check/?q=Cardiac%20Assist%2C%20Inc) [Browse related products](https://isitonrecall.com/products/) [Review source dates](https://isitonrecall.com/sources/)